Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20
Tip Assembly Using Disposable Torque
Wrench
Refer to the CUSA Clarity Ultrasonic Surgical Aspirator
System Operator's Manual and the specific tip IFU
for complete handpiece assembly and disassembly
instructions.
For this procedure, use the CUSA Clarity sterilizable
torque base.
1. Thread a tip onto the handpiece connecting body
by turning the tip clockwise until it is finger tight.
Place the torque base on a flat surface or in the
designated spot in the sterilization tray.
WARNING
Before use, sterilize the torque base in the sterilization
tray with the handpiece.
DRAFT FOR
2. Place the handpiece on the torque base as shown
and make sure that the 2 flat metal sides fit snugly
in the metal slot at the end of the base.
VALIDATION
3. Hold the handpiece in the base and place the
provided torque wrench (color side towards
handpiece) over the tip.
CAUTION
To avoid product damage, always use the torque base
to hold the handpiece while using the torque wrench to
tighten or loosen the tip. Avoid overtorquing the tip.
4. Rotate the torque wrench clockwise until it clicks
twice.
0
NOTE:
Keep the torque wrench and base in the sterile field in
case you need to change a tip during the procedure.
5. Remove the torque wrench from the tip.
6. Remove the handpiece from the torque base.
Completing Handpiece Assembly
Refer to the CUSA Clarity Ultrasonic Surgical Aspirator
System Operator's Manual for instructions on how to
attach the nosecone, flue, and tubing to the handpiece.
Product Information Disclosure
INTEGRA LIFESCIENCES CORPORATION ("INTEGRA")
HAS EXERCISED REASONABLE CARE IN THE
SELECTION OF MATERIALS AND THE MANUFACTURE
OF THESE PRODUCTS. INTEGRA EXCLUDES
ALL WARRANTIES, WHETHER EXPRESSED OR
IMPLIED, INCLUDING BUT NOT LIMITED TO, ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. INTEGRA
SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR
CONSEQUENTIAL LOSS, DAMAGE, OR EXPENSE,
DIRECTLY OR INDIRECTLY ARISING FROM USE OF
THESE PRODUCTS. INTEGRA NEITHER ASSUMES NOR
AUTHORIZES ANY PERSON TO ASSUME FOR IT ANY
OTHER OR ADDITIONAL LIABILITY OR RESPONSIBILITY
IN CONNECTION WITH THESE PRODUCTS.
CAUTION
Federal (USA) law restricts this device to sale by or on
the order of a physician.
Sterility is guaranteed unless the package is opened or
damaged.
0086
Manufacturer
Integra LifeSciences (Ireland) Limited
IDA Business & Technology Park
Sragh
Tullamore
County Offaly
Ireland
USA & Canada: (800) 997-4868
Outside USA: (609) 936 5400
Assembled in: USA
www.integralife.com
U.S. Patent 6,499,358
Integra, the Integra logo and CUSA are registered trademarks of
Integra LifeSciences Corporation or its subsidiaries in the United
States and/or other countries.
©2018 Integra LifeSciences Corporation. All Rights Reserved.
92905544 Rev A 03/18 0924415-1
3
EN – ENGLISH

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis