11. Technical specifications

Model no.
Type
Dimensions (W, H, D)
Weight
Light
Output
Light intensity
Radiation
Operating conditions
Transport and storage conditions
Product classification
Colour temperature of the LEDs
Threshold of short wavelength
Expected service life of the device
Maximum radiance output of the TL 41 Touch
Radiance output
E
: Eye UV-A
UVA
ES: Actinic UV skin & eye
EIR: Infrared radiation hazard exposure
limits for the eye
LIR: Retinal thermal
(mild visual irritation)
LB: Blue light
LR: Retinal thermal
Subject to technical changes.
The serial number is located on the device or in the battery compartment.
Brightness: 10,000 lux (This point about the light intensity is merely for the purpose of information. With regard to
the standard IEC 60601-2-57, this light source is classified as an Exempt Group).
Notes on electromagnetic compatibility
The device complies with the EU Medical Devices Directive 93/42/EEC, the German Medical Devices Act
(Medizinproduktgesetz) and the European standard EN 60601-1-2 (in accordance with IEC61000-3-2,
IEC61000-3-3, IEC61000-4-2, IEC61000-4-3, IEC61000-4-4, IEC61000-4-5, IEC61000-4-6, IEC61000-4-8,
IEC61000-4-11) and is subject to particular precautions with regard to electromagnetic compatibility.
• The device is suitable for use in all environments listed in these instructions for use, including domestic
environments.
• The use of the device may be limited in the presence of electromagnetic disturbances. This could result in
issues such as error messages or the failure of the display/device.
GCE502
TL 41 Touch
270 x 207 x 80 mm
Approx. 420 g
LED
26 watts
10,000 lux (at a distance of approx. 20 cm)
Radiance output outside of the visible range (infrared and UV) is low
enough that no eye or skin damage is anticipated.
0 °C to +35 °C, 15 – 90% relative humidity,
700 – 1060 hPa ambient pressure
-20 °C to +60 °C, 15 – 90% relative humidity,
700 – 1060 hPa ambient pressure
External power supply, protection class II, IP21
6500 Kelvin
???
nm
10,000 hours
Risk group classified
in accordance with
IEC 60601-2-57:2011
Exempt Group
Exempt Group
Exempt Group
Exempt Group
Exempt Group
Exempt Group
15
Maximum value
???
W·m
-2
???
W·m
-2
???
W·m
-2
-2
-1
???
W·m
·sr
???
W·m
-2
·sr
-1
???
W·m
-2
·sr
-1