Transport & Storage Conditions
Storage
Relative Humidity Range
Classifi cation and Approvals
Class I Equipment
Medical equipment with respect to electric shock, fi re, and mechanical hazards only in
accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1.
IPX1 Water Ingress Protection
Continuous Operation
Compliance
FCC Regulations
IC Regulations
Note: Please contact your local Stryker Endoscopy sales representative for information on changes
and new products.

Electromagnetic Compatibility

Like other electrical medical equipment, the WiSe™ 19" HDTV Surgical Display requires special
precautions to ensure electromagnetic compatibility with other electrical medical devices.
To ensure electromagnetic compatibility (EMC), the display must be installed and operated
according to the EMC information provided in this manual. Th e display has been designed and
tested to comply with IEC 60601-1-2 requirements for EMC with other devices.
Warning
Caution
Guidance and Manufacturer's Declaration: Electromagnetic Emissions
Th e WiSe™ 19" HDTV Surgical Display is intended for use in the electromagnetic environment specifi ed below. Th e customer or
the user of the WiSe™ 19" HDTV Surgical Display should ensure it is used in such an environment.
Emissions test
24
-0.4 – 140°F (-18 – 60°C)
15 – 90%
FCC Part 15 Class B
FCC Identifi er: QVXAMM190WTDSW (display)
FCC Part 15 Class C
FCC Identifi er: VQSAMN12100R44 (radio module)
IC: 7680A-AMN12100 (WiSe 19" and 26" HDTV
Surgical Display)
When this device is connected with other electrical equipment, leakage
currents may be additive. To minimize total leakage current per patient,
ensure that all systems are installed according to the requirements of IEC
60601-1-1.
Portable and mobile RF communications equipment may aff ect the normal
function of the display.
Do not use cables or accessories other than those provided with the display,
as this may result in increased electromagnetic emissions or decreased
immunity to such emissions.
If the display is used adjacent to or stacked with other equipment, observe
and verify normal operation of the display in the confi guration in which
it will be used prior to using it in a surgical procedure. Consult the tables
below for guidance in placing the display.