Electromagnetic Compatibility

Like other electrical medical equipment, the VISION ELECT HDTV
(model 240-030-960) monitor requires special precautions to ensure
electromagnetic compatibility with other electrical medical devices. To
ensure electromagnetic compatibility (EMC), the VISION ELECT HDTV
(model 240-030-960) monitor must be installed and operated according to
the EMC information provided in this manual.
Note
The VISION ELECT HDTV (model 240-030-960) monitor
has been designed and tested to comply with IEC
60601-1-2:2003 requirements for EMC with other devices.
Caution
The VISION ELECT HDTV (model 240-030-960) monitor
may be interfered with by other equipment, including
portable and mobile RF communication equipment,
even if such equipment meets the applicable
emissions requirements.
Warning
Using cables or accessories other than those provided
with the VISION ELECT HDTV (model 240-030-960)
monitor, may result in increased electromagnetic
emissions or decreased immunity to such emissions.
Warning
If the VISION ELECT HDTV (model 240-030-960)
monitor is used adjacent to or stacked with other
equipment, observe and verify normal operation of the
VISION ELECT HDTV (model 240-030-960) monitor in
the configuration in which it will be used prior to using
it in a surgical procedure. Consult the tables below for
guidance in placing the VISION ELECT HDTV (model
240-030-960) monitor.
Warning
When this device is connected with other electrical
equipment, leakage currents may be additive. To
minimize total leakage current per patient, ensure that
all systems are installed according to the requirements
of IEC 60601-1-1.