Herunterladen Inhalt Inhalt Diese Seite drucken

Medtronic RestoreULTRA 37712 Implantationsanleitung Seite 8

Multiprogrammierbarer wiederaufladbarer neurostimulator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 31
Filename Date Time
UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Table 1. Operating values for the Model 37712 RestoreULTRA neurostimulator
Programmable parameter Operating range and resolution
Number of defined groups
Number of programs per
group
Electrode configuration
Amplitude
Amplitude – upper patient
limit
Amplitude – lower patient
limit
Pulse width
Pulse width – upper patient
limit
Pulse width – lower patient
limit
Rate
Rate – upper patient limit
Rate – lower patient limit
SoftStart/Stop
Cycling
Scheduled therapy
Interlocks will prevent the use of some parameter combinations.
a
b
No more than 16 programs may be defined within the 8 groups.
c
Only 0.1 V resolution possible from 10.0 V to 10.5 V amplitude.
d
Rate limited to 600 Hz when two programs/group are active, 400 Hz when three programs/group are active
and 300 Hz when four programs/group are active.
6 English  37712
2007-9
b
1 to 8
b
1 to 4
2 to 16 electrodes as anode, cathode, or OFF
0 to 10.5 V with 0.05 V- or 0.1 V-resolution
Tracking limit: programmed value +0 to +4 V (0.5 V-resolution)
Custom limit: programmed value to 10.5 V (same resolution as
amplitude)
Custom limit: 0 V to the programmed value (same resolution as
amplitude)
60 to 1000 µs (10-µs resolution)
Tracking limit: programmed value +0 to +300 µs (60-µs
resolution)
Custom limit: programmed value to 1000 µs (10-µs resolution)
Custom limit: 60 µs to the programmed value (10-µs resolution)
2 to 1200 Hz (resolution: 1 Hz from 2 Hz to 10 Hz, 5 Hz from 10 Hz
to 250 Hz, 10 Hz from 250 Hz to 500 Hz, 20 Hz from 500 Hz to
1000 Hz, 50 Hz from 1000 Hz to 1200 Hz)
Tracking limit: programmed value +0, +10, +20, +50, +100 Hz
Custom limit: programmed value to 1200 Hz (same resolution as
rate)
Custom limit: 2 to the programmed value (same resolution as
rate)
OFF, ON: 1, 2, 4, or 8 second ramp duration
OFF, ON: 0.1 s to 30 min (resolution: 0.1 s from 0.1 s to 1 s, 1 s
from 1 s to 1 min, 1 min from 1 min to 30 min)
OFF, ON: 1 to 8 events per 24-h period
M927169A004
Medtronic Confidential
IPGTemplate.fm
Version: 10-23-2007
a
c
d

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis