Description

The Medtronic Model 37712 RestoreULTRA Neurostimulator is part of a neurostimulation
system for pain therapy.

Package contents

Neurostimulator
Torque wrench
Product literature
Warranty card (USA only)
Registration form

Patient identification card

Patient identification card
A patient identification card is packaged with this device. Advise the patient to carry the
identification card at all times.
The patient identification card packaged with the device is temporary; a permanent card
! USA
will be mailed to the patient when Medtronic receives the registration form.
The implant registration form registers the device warranties and creates a record of the device
in Medtronic's implant data system.

Device specifications

The neurostimulator is a multi-programmable, rechargeable device that delivers stimulation
through 1 or more leads. The stimulation settings are stored in programs. A program is a
specific combination of pulse width, rate, and amplitude settings acting on a specific electrode
combination (up to 16 electrodes per program). Up to four programs can be combined into a
group. When using more than one program, the pulses are delivered sequentially—first a
pulse from one program, then a pulse from the next program.
Pulse width, amplitude, and electrode polarity for each program within the group can have
different values. Rate, rate limits, ramping, and cycling for each program within the group have
the same values.
M927169A004
Medtronic Confidential
IPGTemplate.fm
Version: 10-23-2007
37712 English 5
2007-9