SAFETY STANDArDS

14
- The dental lamp MAIA is designed to light up the oral cavity of the patient
during dental procedures.
- The lamp must be used by qualified personnel.
The lamp must only be installed by specialised personnel.
Ensure that power-line voltage indicated on the information plate corresponds
to that of the supply network.
The dental lamp must be installed on a specific control and power feed device,
such as dental chairs, or connected to a wiring system that meets IEC 364-1
standards and "national rules for installation of wiring systems in premises
allocated for medical use."
The lamp must be installed with a multipolar device to separate it from the supply
network. Said device shall meet IEC/EN 61058 standards. A green status light
shall be inserted to indicate that the lamp is powered.
- Do not do any maintenance work on the lamp when the power supply is on or
in the presence of the patient. Disconnect the power cable from the supply socket
before starting work.
- Do not insert any objects or equipment that could come in contact with live spots
into the slits of the lamp head.
- The articulated arm and hinge joints of the head allow the light beam to be
correctly positioned. Do not overload the arms and hinge joints with impact at
the end of stroke.
- Do not use a fixed light beam for patients at risk (e.g. children, adults with eye
disorders). Always use appropriate protection devices and precautions. Faro
suggests using BLUE-BAN or BABY BLUE-BAN (for children) protective
eyeware.
Faro eyeware are individual protection devices for protecting the eyes of the patient
from the potential photobiological risks of light. Faro recommends using them
for patients who are particularly exposed to such risks, namely children and adults
with eye disorders who are taking photosensitive substances. The eyeware must
be worn for the entire duration of treatment. They shall be cleaned and disinfected
before and after use to prevent cross-contamination. Sterilise only at 121°C. Any
scratches and/or haloes do not compromise their efficacy. Do not use in case of
mechanical breakage.
- Not suitable for installation in rooms containing flammable gas.
- Do not spray detergents-disinfectants directly on the lamp head.
- The information plate with LED indications is fixed to the rear arm.
- Any inadequate performance of the device will not impair patient safety in any way.
- For cleaning plastic parts of the lamp MAIA, DO NOT uSE detergents-
disinfectants containing: AMMONIuM HYDrOXIDE SODIuM HYDrOXIDE
METHYlENE CHlOrIDE METHYl AlCOHOl.
Failure to comply with this indication might cause: risk of breakage of plastic
parts • risk of structural breakdown of hinge joints with potential falling of the
articulated arm. In case of doubt, please contact Customer Care Faro.
- Do not perform servicing or replacement operations on parts other than those
specified in the manual. Any intervention not indicated in the same might impair
the safety features of the device.
- For disinfection of surfaces, use hydro-alcohol-based disinfectants.
- Do not leave small components of the device unattended or at the reach of exposed