TrOUBLESHOOTING

The lamp will not switch on
Intensity has
considerably diminished
Handles will not snap on
or they snap off with difficulty
Patches have appeared
on the reflectors (dishes)
PErIODIc cHEcKS
 Check the lack of smooth motion in hinge joints of the arms (yearly)
 Check readability of information on the plate (yearly)
 Power safety checks: (every two years)
1. Rigidity
2. Dispersion.
 Light checks: (every five years or 10,000 hours of function)
1. Maximum irradiance: >35,000 lux
2. Blue light on emitted spectrum measured in W/m
If any anomalies occur during the periodic check, immediately contact
Customer Care faro.
AcOUSTIc SIGNALS
MAX
= 2 beep only with prox
MIN
= 1 beep
1 Beep = at commands
1 Beep = on switching on
OpL** = 1 beep per sec. for 30 seconds
OpL**: Load LED disconnected
– Check that it is connected to the power supply.
– Check status of fuses.
– If none of these causes applies, contact Technical
Support.
– Clean the dishes.
In case of failure to restore initial conditions of
irradiance, contact Technical Support.
– Check that the position of the locking screw on the
handle is completely open.
– Clean the surfaces with the specific product
"Faro Perflex".
– Clean the surfaces with isopropyl alcohol.
– An unsuitable product for cleaning or disinfection
has been used, thus damaging the surfaces.
– Contact Customer Care FARO.
: <100
2
23 23