EN-6
RS232 OUTPUT
With the interface installed the balance will operate in several ways according to the
settings.
Print format can be switched by xFMT user command (see command table in
RS232 INPUT section).
New Scout Print Format (Default Format):
Output String (Non Check Weighing Applications):
[weight]
[space]
[unit]
[space]
[stability indicator]
[space]
[T/N/G/PT]
[Term]
Note: All of the fields have fixed length.
Output example:
*****192.21_****g_*_**
*******0.01_****g_?_**
*******95.0_****g_*_*N
******169.6_****g_*_*G
*******95.0_****g_*_*N
*******74.6_****g_*_*T
Output String (Check Weighing Application):
[weight]
[space]
[unit]
[space]
[stability indicator]
[space]
[T/N/G/PT]
[space]
[application status]
[Term]
Example:
11 characters (right justified)
1 character
5 characters (right justified)
1 character
1 character; "?" when unstable, space when stable
1 character
2 characters (right justified)
2 characters
A.Print: off; Stable: off
A.Print: off; Stable: on
11 characters (right justified)
1 character
5 characters (right justified)
1 character
1 character; "?" when unstable, space when stable
1 character
2 characters (right justified)
1 character
6 characters (right justified)
2 characters
Stable reading
unstable reading
Content / Result -> on
Content / Gross -> on
Content / Net -> on
Content / Tare -> on
RS232 Interface