For PJX balances:
Command
Function
IP
Immediate Print of displayed weight (stable or unstable).
P
Print displayed weight (stable or unstable).
CP
Continuous Print.
SP
Print on Stability.
SLP
Set Auto Print to On Stability, allow non-zero displayed weight be printed.
SLZP
Set Auto Print to On Stability, allow both stable non-zero weight and stable zero reading printed.
xP
Set Auto Print to Interval Print, x = print interval (1-3600 sec), 0P disable the interval Print
0P
0P disable interval print, continuous print or print on stability
H
Enter Print Header Lines, the format is: H x "header string".
Where x = line number 1 to 3, "header string" can be up to 24 alphanumeric characters.
If no string in the command, "H x" will read the stored header x.
Z
Same as pressing Zero Key.
T
Same as pressing Tare Key.
xT
Establish a preset Tare value in displayed unit. x = preset tare value. Sending 0T clears tare (if allowed).
PT
Prints Tare weight stored in memory.
PM
Print current application mode (weighing mode).
xM
Set current application mode to x. x depends on applications
1 – Weigh
2 – Count
3 – Percent
4 – Density
M
Scroll to the next enabled mode.
PU
Print Current weighing unit: g, kg, lb, oz, etc....
xU
Set balance to unit x: g, kg etc.
1 – g
2 – kg
3 – mg
4 – ct
...
U
Scroll to the next enabled unit.
ON
Brings out of Standby
OFF
Goes to Standby.
C
Begin Span Calibration
IC
Begin Internal Calibration.
AC
Abort Calibration.
PSN
Print Serial Number.
PV
Print Version: print name, software revision and LFT ON (if LFT is set ON).
x#
Set Counting APW (x) in grams. (must have APW stored)
P#
Print Counting application APW.
x%
Set Percent application reference weight (x) in grams. (must have reference weight stored)
P%
Print Percent application reference weight.
PTIME
Print current time.
PDATE
Print current date.
xTIME
Set Time x format: hh:mm:ss
xDATE
Set Date x format: mm/dd/yyyy
xS
0 = print unstable data, 1 = print stable only
xRL
Enable or disable OK response to non-print commands: x=0 to disable, x=1 to enable.
xT
Pre-tare the container weight (x) in grams.
RS232 Interface