Installation Manual
When using PR 6041/80S adaptor plates (13):
Place the adaptor plate (13) onto the foundation (8) within the outline marking (see page 18).
Mount the adaptor plate using bolts (13). It is essential to observe the property classes and
tightening torques of nuts (11) and washers (12); refer to Chapter 4.1.2.
Insert the O-ring (14) into the groove of the lower plate (15).
Lower the vessel (19) to the foundation (8)/adaptor plates (13) and mount the lower plate (15) using
bolts. It is essential to observe the property classes and tightening torques of nuts (11) and
washers (12); refer to Chapter 4.1.2.
Only applicable to PR 6241/12...52 load cells:
Take the two bushes out of the bag and fit them on the two pins (see page18).
Remove the auxiliary mounting plates (18) (if necessary, unload them using the jack-up).
Release the locknut (2), see Chapter 2.3.3.
Turn the bolt (1) of the lift-off protector at the hex (3) down slowly to lower the vessel to the load cell
and adjust the internal lift-off protector (4), see Chapter 2.3.3.
Check whether pivot (9) or/and roller (10) can move freely.
If pivot (9) or/and roller (10) cannot be moved, the weighing module is distorted and must be adjusted:
Release the bolts (16).
Shift the stopper (17), until pivot (9) or/and roller (10) can move freely again.
Re-tighten the bolts (16). Make sure to use the correct tightening torques, see Chapter 4.1.2.
Shift off the bellow from the vessel foot.
Fit the bellow, press it firmly into the groove of the upper and lower plate circumferentially, and make
sure that the transitions are flush.