  • DEUTSCH, seite 31
Closed to Atmosphere: A minimum of 3 cm H
that the anti-asphyxia valve closes.
Open to Atmosphere: In the absence of positive airway pressure, the anti-
asphyxia valve will open.
Inspiratory and Expiratory Resistance with Anti-Asphyxia Valve Open to
Atmosphere at 50 SLPM
Inspiratory Resistance
Expiratory Resistance
Deadspace
S
150.4 mL
M
148.8 mL
L
167.8 mL
Sound Levels
A-weighted Sound Power Level
A-weighted Sound Pressure Level at 1m 24.2 dBA
Disposal
Dispose of in accordance with local regulations.

Storage Conditions

Temperature: -4° F to 140° F (-20° C to 60° C)
Relative Humidity: 15% to 95%, non-condensing

LIMITED WARRANTY

Respironics, Inc. warrants that its mask systems (including mask frame and cushion)
(the "Product") shall be free from defects of workmanship and materials for a period
of ninety (90) days from the date of purchase (the "Warranty Period").
If the Product fails under normal conditions of use during the Warranty Period and
the Product is returned to Respironics within the Warranty Period, Respironics will
replace the Product. This warranty is nontransferable and only applies to the original
owner of the Product. The foregoing replacement remedy will be the sole remedy
for breach of the foregoing warranty.
This warranty does not cover damage caused by accident, misuse, abuse,
negligence, alteration, failure to use or maintain the Product under conditions of
normal use and in accordance with the terms of the product literature, and other
defects not related to materials or workmanship.