Connector
1
pH/mV
Figure 4
1 BNC sensor connector
2 Temperature compensation sensor
connector
8
2
Temp

Operational Procedures

Buffer Table Selection
There are two sets of buffer available
for this meter: 4.00, 7.00, 10.00 pH and
4.01, 7.00, 6.86, 9.18 pH.
Press the mode key to toggle between
pH and mV modes. In pH mode, press
the setup key to go into Buffer
Selection Mode.
Buffer Selection Mode
Use [
] and [
] keys to select buffer
set. Press the enter key to confirm
buffer set and to exit Buffer Selection
Mode.