7.2 Troubleshooting

No indicator lights on when compressor turned On.
RED Light On
Compressor cycles on and Off
or
O
Below Normal
2
(yellow) and filling (green) lights are alternating
or
O
below normal (yellow) light goes On and Off
2
1145804-F-00
Problem
1.
Turn compressor Off (O).
2.
Check that power cord is properly plugged into compressor and
wall outlet.
3.
Turn the compressor On ( I ).
4.
Clean air filters. Refer to Cleaning the Compressor Filter in the
Maintenance section of the manual.
5.
If performing Steps 1-4 does not cause any indicator lights to come
on, contact your home health care provider or Invacare for service.
1.
Check the cylinder gauge to see if it is full:
Cylinder full - Remove the cylinder and turn the unit Off.
Perform the Cylinder Prefill Inspection in the Usage section of
the manual with a cylinder that is not full.
Cylinder not full - Proceed to Step 2.
2.
Check that the cylinder is connected properly.
Turn the compressor Off (O).
Remove the cylinder and reinstall.
Turn the compressor On ( I ). If cylinder is not full and is
connected properly, the compressor may have internal failure.
Contact your home health care provider or Invacare.
1.
Check the concentrator liter flow setting is at 2.5 L/min or less
for Platinum 5 or Perfecto
Platinum 9 concentrators.
If necessary, adjust the liter flow setting (refer to the concentrator
User manual). Wait at least 20 minutes for the GREEN light to
appear. If it does not appear, proceed to step 2.
2.
The concentrator may need service, contact your home health
care provider or Invacare.
Troubleshooting
Solution
concentrators or 5 L/min or less for
2
29