EC RTD
9
Technical data
The technical data of customised versions may differ from the data in these instructions.
Please observe the information specified on the type plate.
Applicable to a temperature range from 18 to 28 °C, within one year after calibration.
9.1 Characteristics EC RTD
Characteristics
Output
Resistance (Ohm)
- Output range
- Resolution
- Accuracy
- Remarks
RTD (Pt100)
- Output range
- Resolution
- Accuracy
- Remarks
RTD (Cu50)
- Output range
- Resolution
- Accuracy
- Remarks
Note 1: without accessory lead resistance
Note 2: range of exciting current: 0.5mA to 2mA; max. Output voltage: ≤2V
Technical changes reserved
EC RTD
0.0...400.0 Ω
0.1 Ω
± (0.05 % of set value + 0.2 Ω)
±1 mA exciting current
Notes 1 & 2
-200.0...850.0 °C
0.1 °C
± (0.05 % of set value + 0.6 °C)
±1 mA exciting current, use Pt100-385
Measuring current 1 mA
Notes 1 & 2
-50...150.0 °C
0.1 °C
± (0.05 % of set value + 0.6 °C)
±1 mA exciting current
Notes 1 & 2
Technical data