Technical Data
PERFUSOFP
secura
FT
Nominal voltage (V =-=-=)
Nominal current (A)
Classilication
AFI:
Moisture protection:
Duration of operation:
11.0 ... 16.0
0.35
according to IEC 601-1 (Identical to DIN IEC 601 Teil
1NDE 0750 Teil 1 and
es
5724 part 1)
Protection class:
11,
Type: CF
B (according to DIN VDE-0871 and DIN VDE 0875,
respectively EEC-Directive 82/499 and CISPR 11)
drip-proof
100 % (cont)
Environmentsi conditions (without condensation)
a. operation
- relative humidity:
- temperature:
- atm. pressure:
b. Transport and storage
- relative humidity:
- temperature:
Accumulator
type:
Accumulator:
Aecharge time:
Pre-alarm:
Deviation of preselected
delivery rate (including the
disposables recornmended by
the manufacturer:
Automatical alarm activated in
case
01
dosage failure:
Occlusion pressure:
Delivery range:
Preselection of volume:
Staft call:
Weight:
Dimensions (Ixwxd) in mm:
30%
75%
+5°C
+4O°C
500
mbar ... 1060 moor
10%
90%
-25°C
+55°C
NiCad, Ord.-Nr. 3317024/0 (rechargeable)
>
4
h
at max. delivery rate
>
16 h
3 min before end of infusion
±
2
%,
duration of measurement
>
1
h,
delivered
.volume
>
2 ml
At malfunction of the unit the pump will stop in ease of
dosage failure of max. 0.2
J.lI.
<
120 kPa
«
1.2 bar)
0.1
99.9
mllh (inerements
01
0.1 mllh)
0.1
99.9
ml; 100 ...
999
ml
±
1 ml
max. 24 VI1 A/24 VA pol. connect. interehangeable
ca. 2.2 kg
310x155x155
Ord.-No.
VoItageICurrentrequenz
-
.
Currenl
PIug
....
,
382386813
3823875/6 PIug-in
3823876/4 power
suppIy
38238nl2
87180312
345031110 integr.
345031012 power
suppIy
220 V ­
SOI6O
Hz
220
V -
SO/50 Hz
240
V ­
SO/50 Hz
220 V -
SOI6O Hz
120
V ­
50/60
Hz
100/1101120 V -
50/60
Hz
20012301240 V -
SO/50
Hz
O,03A
O,03A
O,03A
O,03A
O,06A
O,09A
O,04A
(SChuico)
(Euro)
•..
(UK)
(UK)
(Hospital Grade)