Performance
Flow rate
Timing
Positive Pressure
Safety Relief
Valve:
Gauge Accuracy
Adult
Pediatric
Neonatal
Physical
Dimensions
Weight
Battery
Lithium
full scale deflection
1
full range at 22°C
2
not adjustable, without fittings at full increase
3
Timing is preset at factory using a source vacuum level of 500 mmHg (66.7 kPa). Timing may vary in relation
4
to the source vacuum pressure.

ENVIRONMENTAL SPECIFICATIONS

Operating Temperature Range
Storage Temperature Range
Operating and storage Relative Humidity
Ingress Protection Ratings:

Specifications

Intermittent
Continuous
15 seconds ON, 8 seconds OFF (± 3 seconds)
Located in the vacuum supply line to prevent pressurization of
the patient connection by failed injector vacuum (Venturi) units, or
inadvertent cross connection to pressured gases.
0-200 mmHg (0-26.7 KPA)
0-160 mmHg (0-21.3 KPA)
0-160 mmHg (0-21.3 KPA)
0-100 mmHg (0-13.3 KPA)
6.5"H x 2.8"W x 4.8"D
(16.5 cm x 7.1 cm x 12.2 cm)
20 oz (567 grams)
Two 2/3 AA batteries, 3.6V, 1.6 Ah, lithium
50 to 104ºF (10 to 40ºC)
-13 to 158ºF (-25 to 70ºC)
5 to 95% RH (Non-condensing)
IP 20
8700-0009-000
Adult
Pediatric
8 LPM
80 LPM
80 LPM
3
0-16 LPM (Preset at 8 LPM)
Can be adjustable
Starts in the ON cycle
Analog
1
±5% Full-scale
±5% Full-scale
±5% Full-scale
-
(Rev. 12) 01/2017
Neonatal
8 LPM
8 LPM
80 LPM
3
3
4
Digital
±1% Full-scale
±1% Full-scale
-
±1% Full-scale
2
5