EN
Performance
Flow rate
1
Positive Pressure
Safety Relief Valve
Gauge Accuracy
Standard
High
Pediatric
Neonatal
C.A.S.S.
Physical
Dimensions
Weight
Battery
Lithium
without fittings at full increase setting depending on supply vacuum and open air flow.
1
full scale deflection
2
full range at 22°C
3

ENVIRONMENTAL SPECIFICATIONS

Operating Temperature Range
Storage and Transport Temperature Range
Operating Atmospheric Pressure Range
Storage and Transport Atmospheric Pressure Range
Operating and Storage Relative Humidity
Ingress Protection Rating

Specifications

Standard
High
80 L/min
80 L/min
Located in the vacuum supply line to prevent pressurization of
the patient connection by failed injector vacuum (Venturi) units, or
inadvertent cross connection to pressured gases.
Gauge Range
0-200 mmHg (0-26.7 kPa)
0-760 mmHg (0-101.3 kPa)
0-160 mmHg (0-21.3 kPa)
0-160 mmHg (0-21.3 kPa)
0-100 mmHg (0-13.3 kPa)
0-160 mmHg (0-21.3 kPa)
6.7"H x 2.8"W x 4.8"D
(17.02 cm x 7.1 cm x 12.2 cm)
1 lb 4 oz (0.57 kg)
Two 2/3 AA batteries, 3.6V, 1.6 Ah, lithium
8700-0007-000 (Rev. 16) 10/2020
Pediatric
Neonatal
80 L/min
80 L/min
Analog
2
±5% Full-scale
±5% Full-scale
±5% Full-scale
±5% Full-scale
±5% Full-scale
±5% Full-scale
50 to 104ºF (10 to 40ºC)
-13 to 158ºF (-25 to 70ºC)
80 to 106 kPa
50 to 106 kPa
5 to 95% RH (Non-condensing)
IP 20
C.A.S.S.
80 L/min
Digital
3
±1% Full-scale
±1% Full-scale
±1% Full-scale
±1% Full-scale
±1% Full-scale
±1% Full-scale
5