Cleaning and Disinfection

Cleaning and Disinfection
Routine cleaning of the regulator is recommended as a standard procedure after each use.
Wipe all exterior surfaces with a solution of water and mild detergent and/or an approved cleaning solution.
Should misuse occur resulting in accidental flooding of the regulator, the regulator may be sterilized after
cleaning using ethylene oxide (ETO). See Section 4.2 (Sterilization) of the PTS-ISU service manual. After
sterilization follow the service checkout procedures in Section 7.0 (Service Checkout Procedure) of the PTS-
ISU service manual.
Approved Cleaning Solutions
Sodium Hypochlorite 0.5% (Bleach): Mixture of 13 fl. oz. of bleach to 1 gallon (128 fl. oz.) tap water
Isopropyl Alcohol 70%
Hydrogen Peroxide 3%
Cavicide
Ready to use full strength
®
WARNING
After patient use, regulators may be contaminated. Handle in accordance with your
hospital's infection control policy.

Troubleshooting

If the regulator does not operate and you have performed the Pre-Use Checkout Procedure, remove the
regulator from service and refer to PTS-ISU service manual or refer servicing to qualified service personnel.
Timing Cycle Adjustment
For timing cycle adjustment, refer to PTS-ISU service manual, section 7.
Installation Procedure for Adapters/Probes and Fittings
For adapter/probe installation, refer to PTS-ISU service manual, section 8.6.