Knick Portavo 902 PH Kurzübersicht Seite 17

Vorschau ausblenden Andere Handbücher für Portavo 902 PH:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 21
Specifications
Sensor standardization
Operating modes
*)
Calimatic buffer sets
*)
Permissible calibration range
Calibration timer
*)
Sensoface
Evaluation of
*) User-defined
pH calibration
*)
CALIMATIC
MANUAL
DATA INPUT
-01- Mettler-Toledo
-02- Knick CaliMat
-03- Ciba (94)
-04- NIST technical
-05- NIST standard
-06- HACH
-07- WTW techn. buffers 2.00/4.01/7.00/10.00
-08- Hamilton
-09- Reagecon
-10- DIN 19267
Zero point
With ISFET:
Operating point
(asymmetry)
Slope
(possibly restricting notes from Sensoface)
Interval 1 ... 99 days, can be switched off
Provides information on the sensor condition
zero/slope, response, calibration interval
Calibration with automatic buffer recognition
Manual calibration with entry of individual
buffer values
Data entry of zero and slope
2.00/4.01/7.00/9.21
2.00/4.00/7.00/9.00/12.00
2.06/4.00/7.00/10.00
1.68/4.00/7.00/10.01/12.46
1.679/4.006/6.865/9.180
4.01/7.00/10.01 /12.00
2.00/4.01/7.00/10.01/12.00
2.00/4.00/7.00/9.00/12.00
1.09/4.65/6.79/9.23/12.75
pH 6 ... 8
-750 ... +750 mV
approx. 74 ... 104 %
17