Herunterladen Inhalt Inhalt Diese Seite drucken

Anhang 1: Emv-Tabellen; Appendice 1: Tabelle Di Compatibilità Elettromagnetica; Bijlage 1: Emc-Tabellen; Dodatek 1: Tabele Kompatybilności Elektromagnetycznej - Aerogen USB Controller Bedienungsanleitung

Usb-steuereinheitssystem
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Appendix 1: EMC Tables
The following tables are provided in accordance with IEC/EN 60601-1-2:
Table 12.
Guidance and manufacturer's declaration – electromagnetic emissions
The Aerogen USB Controller system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen USB Controller
system should assure that it is used in such an environment.
Emissions test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations /
flicker emissions
IEC/EN 61000-3-3
Compliance
Electromagnetic Environment -
Guidance
Group 1
The Aerogen USB Controller system uses
RF energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interference
in nearby electronic equipment.
Class B
The Aerogen USB Controller system is
suitable for use in all establishments,
including domestic establishments and
Not
those directly connected to the public
Applicable
low voltage power supply network that
Not
supplies buildings used for domestic
Applicable
purposes.
Aerogen
®