Maintenance

9.1

Cleaning

Only clean the outside of the plasma generator.
Ensure that the plasma generator is disconnected from the power supply.
Only clean the plasma generator with a cloth dampened in water. Do not use solvents
to clean the plasma generator!
9.2

Changing the nozzle

Proceed as follows:
Ensure that the plasma generator is disconnected from the power supply.
Press the catch buttons on both sides of the nozzle insert.
Pull the nozzle insert out of the housing.
Put the new nozzle insert in the housing. If necessary, press the catch buttons initially
and do not angle the insert.
Apply a little pressure and slide the nozzle insert in until it engages in the lock position.
Attention – Damage to device!
Never touch the piezo crystal on the front side of the device with sharp-edged objects.
This component may be damaged by improper handling.