AILY
• Check and note the cool-room temperature using the thermometer.
If the temperature does not correspond to the level required, rectify using the thermostat button. Should this fail,
refer to chapter entitled "Fault Finding".
• Store the vaccines away in their correct order to enable air to circulate freely between the individual
packages.
• Remove any residual liquid from the inner container immediately.
W
EEKLY
• Check for the ice layer on the evaporator. If the ice layer is thicker than 1 cm, the device must be defrosted. (See
section "Defrosting")
• Clean the lid sealing and verify that the lid is locking tightly.
M
ONTHLY
• Use a soft brush to free the unit from dust.
• Clean the device inside and outside. Use lukewarm water with a mild cleaning agent. Abrasive or caustic cleaning
agents, steel wool, scouring sponges, or chemical cleaning agents should not be used under any circumstances.
After the cleaning is finished, the device should be dried properly.
• Clean the lid sealing.
• Check the hinges and lockings for tight fastening and verify that the lid is locking tightly.
A
LL REPAIR AND MAINTENANCE WORK MUST BE PERFORMED
To avoid an unloading of the battery, the electronic unit is fitted with a bat-
tery protection that cuts in and out the compressor, on the voltage values
shown in the table.
The electronic unit of the RCW50DC is fitted with a fuse. In case of a
failure of the appliance, please check this fuse and replace if neces-
sary :
• Unmount the cooling unit cover
• Tear off the fuse.
• The mounting is done in reverse order
A
:
T T E N T I O N
A FAULTY FUSE MIGHT ONLY BE REPLACED BY A FUSE OF
R
OUTINE
BY APPROVED EXPERTS ONLY
B
P
ATTERY
ROTECTION
R
EPLACEMENT OF THE FUSE
.
THE SAME VALUE
M
AINTENANCE
!
(RCW50DC
ONLY
(RCW50DC
- 9 -
)
compressor
cut-off
voltage
12V
10.4V
24V
22.8V
)
ONLY
fuse
cut-in
voltage
11.7 V
24.2 V