NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
a
phones and land mobile radios, amateur radio, AM an FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the device or system is used exceeds the applicable RF
compliance level above, the device or system should be observed to verify normal operation. If
abnormal performance is obsered, additional measures may be necessary, such as reorienting
or relocating the EUTs.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3] V/m.
b
A26391-K270-Z290-1-8N19
Table 3: continued
edition 1
,
CELSIUS H270 MED and Components
English - 7