Table 2: Guidance and manufacturer's declaration – electromagnetic
The device or system is intended for use in the electromagnetic environment specified below. The
customer or the user of the device or system should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
inter-ruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the AC mains voltage prior to application of the test level.
T
A26391-K270-Z290-1-8N19
immunity
IEC 60601
test level
+/- 6 kV contact
+/- 8 kV air
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 1 kV differential
mode
+/- 2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
<40% U
T
(>60% dip in U
)
T
for 5 cycles
<70% U
T
(>30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
edition 1
,
CELSIUS H270 MED and Components
Compliance
Electromagnetic
level
environment - guidance
+/-6 kV
Floors should be wood
+/-8 kV
concrete or ceramic tile. If
floors are covered with
synthetic material the relative
humidity should be at least
30%.
+/- 2 kV
Mains power quality should
+/- 1 kV
be that of a typical
commercial or hospital
environment.
+/- 2 kV
Mains power quality should
+/- 1 kV
be that of a typical
commercial or hospital
environment.
passed, A
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of the
passed, A
device or system requires
continued operation during
power mains interruptions, it
passed, A
is recommended that the
device or system is powered
by an uninterruptible power
supply or a battery.
passed, C
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
English - 5