Herunterladen Inhalt Inhalt Diese Seite drucken

Fujitsu CELSIUS H270 MED Gebrauchsanweisung Seite 22

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 9
CELSIUS H270 MED and Components
Table 3: Guidance and manufacturers declaration – electromagnetic
The device or system is intended for use in the electromagnetic environment specified below. The
customer or the user of the device or system should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
6 - English
immunity
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications
equipment should be used no closer to any part of
the device or system, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
[3] V
3.5
=
d
[3] V/m
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, c
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
P
V
1
80 MHz to 800 MHz
800 MHz to 2.5 GHz
b
A26391-K270-Z290-1-8N19, edition 1
a

Quicklinks ausblenden: