AIH40Ex
EU Declaration of Conformity
Hans Turck GmbH & Co. KG | Witzlebenstraße 7 | 45472 Muelheim an der Ruhr | Germany | Tel. +49 208 4952-0 | Fax +49 208 4952-264 | more@turck.com | www.turck.com
DE
Kurzbetriebsanleitung
Certification Data | Technical Data
Approvals and Markings
Electrical Data
Terminal connection x1 + x4
Approvals
Marking parts in acc. with
(Wiring Diagram I, x = channel no.)
ATEX-directive EN 60079-0/ -11
Measuring transducer
ATEX Certificate
É II 2 (1) G
Ex ib [ia Ga] IIC T4 Gb
circuits for passive sensors
number:
É II (1) D
[Ex ia IIIC Da]
(intrinsically safe acc. to
PTB 00 ATEX 2059X
EN 60079-11 Ex ia IIC or
Ex ia IIIC
Max. output voltage U
IECEx Certification
Ex ib [ia Ga] IIC T4 Gb
Max. output current I
number:
[Ex ia III C Da]
Max. output power P
IECEx PTB 12.0018X
Characteristic
U
Q
Internal resistance R
Internal inductance L
Internal capacitance C
External Inductance C
External capacitance L
EN
Quick-Start Guide
Technical Data
Type designation
Ident no.
Supply voltage
Max. values per channel:
Power consumption
Galvanic isolation
Number of channels
22.1 V
0
Input circuits
93 mA
0
640 mW
0
Supply voltage
trapezoidal
HART® impedance
27.54 V
Overload capability
298 Ω
Low level control
i
Short-circuit
220 µH
i
Wire-break
1.1 nF
i
Ex ia bzw. Ex ib
o
Resolution
o
IIC/IIIC
IIB/IIIC
reference temperature
L
0.5 mH
2 mH
full range
o
rel. Measuring inaccuracy
C
65 nF
270 nF
o
(including linearity, hysteresis
and repeatability)
abs. Measuring inaccuracy
(including linearity, hysteresis
and repeatability)
Linearity deviation
Temperature drift
Rise time/fall time
Max. measurement inaccura-
cy under EMC influence
Connection mode
Protection class
Ambient temperature Tamb
Relative humidity
EMC
AIH40Ex
6884001
via module rack, central
power supply
≤ 3 W
to int. bus and supply circuit
4-channel
0/4...20 mA intrinsically safe
acc. to EN 60079-11
15.1 VDC at 21 mA
> 240 Ω
> 21 mA
< 3.6 mA
> 5 V
< 2 mA (only in live zero
mode)
1 µA
25 °C
0...20 mA
≤ 0.1 % of full range at 25 °C
≤ ± 20 µA at 25 °C
≤ 0.1 % full range
≤ 0.005 % of full range/K
≤ 50 ms (10...90 %)
≤ 0.1 % of full range with
shielded signal cable
≤ 1 % of full range with
unshielded signal cable
module, plugged on rack
IP20
-20...+60 °C
≤ 93 % at 40 °C acc. to EN
60068-2-78
acc. EN 61326-1 (2013)
acc. to Namur NE21 (2012)