Herunterladen Inhalt Inhalt Diese Seite drucken

Dentaurum Polyclav Gebrauchsanweisung Seite 11

Vorschau ausblenden Andere Handbücher für Polyclav:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Dear Customer
Thank you for having chosen a quality product from Dentaurum.
In order to use this product at its best for you and your patients, it is important
to study and follow these directions for use carefully.
The written instructions cannot cover all eventualities during operation.
For this reason our Hotline is available to answer any other questions and
ideas that may arise.
Due to a constant development we recommend, even when you use
the same products frequently, to study the enclosed latest issue of the instructions
for use or refer to our website at www.dentaurum.de.
1. Manufacturer
Dentaurum GmbH & Co. KG
Turnstraße 31 · 75228 Ispringen · Germany
2. With reference to our quality standards
Dentaurum ensures a faultless quality of the products manufactured by us.
These recommendations are based upon our own experiences. The user
himself is responsible for the processing of the products. Responsibility for
failures cannot be taken, as we have no influence on the processing on site.
EC Declaration of Conformity
Dentaurum GmbH & Co. KG · Turnstraße 31 · 75228 Ispringen
hereby declares that the design and construction of the laboratory equipment
described below, including the version marketed by us, comply with the basic
regulations governing safety and health as stated in the EC Guidelines.
This declaration will become invalid if the laboratory equipment is modified or
altered in any way without our prior consent.
Description of the unit:
Type:
EC Guidelines:
The identification plate carries the following information:
Manufacturer:
Maximum pressure:
Operating pressure:
Operating temperature:
Pressure of the pressure release system: max. 8 bar
Capacity:
The serial number and year of manufacture are stamped near the handles of
lid and body of the unit.
The unit is equipped with a class 1.6 manometer having a red line marker at
2.5 bar and a safety valve set at 2.5 bar.
Recurrent testing of pressure vessel must be ordered by user.
We recommend to pressure test the unit at least every 5 years.
The unit must only be operated in accordance with the enclosed
operating instructions.
Date and manufacturer's signature
EN
Polyclav
(REF 070-000-00)
®
Pressure Equipment
for Polymerisation
RL 97/23/EC
Pressure Equipment Directive
Dentaurum Pforzheim
2.5 bar
2.0 ... 2.2 bar
max. 50 °C/122 °F
4 litres
- i.V. Dipl.Ing.(FH) K. Merkle -
Production Manager
 en
polyclav
®
11

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis