4.11. Check internal storage
1. Press "
2. Select "
3. Press "
4. Press "►" or "◄" to select the desired image.
5. Press the "
6. Press the "▲" button and on the screen appears
"Delete photo? Yes / no", and press "
or press "
7. Press to finish writing the "
4.12. Emissivity
The emission factor of an object influences the measurement result
massively and can be adjusted from 0.01 to 1.00. Many common
measurement objects (such as wood, water, skin and textiles) have
a matte surface and high infrared radiation, which is why the
standard emission factor of this device is set to 0.95. For semi-matt
objects, the emission is lower at about 0.85 and at half-glossy
objects even lower, at about 0.6. shiny objects have the lowest
infrared radiation and therefore indicate a wrong reading if the
emission emission setting is wrong. Usually, the infrared radiation at
glossy surfaces is about 0.3.
With an infrared temperature measurement the different surface
qualities must be considered, since these have an influence on the
measurement result. The emission factor describes the
characteristic of the reflectivity of a material. The duller and darker
the object, the more accurate the measurement becomes.