MAINTENANcE

Inspect suction tubing and container for leaks, cracks, etc. before each use.
DANGER
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal components. If service is required, return unit to a
qualified DeVilbiss Healthcare provider or an authorized service center. Opening or tampering with the unit will void warranty.
changing Filter cartridge (Disposable container):
Turn unit
"off".
1.
2. Remove filter cartridge and 4
3.
Install new cartridge and tubing.
NOTE– Do not substitute any other material for this filter cartridge. Substitution may lead to contamination or poor performance; use only DeVilbiss filter cartridges.
NOTE– The filter cartridge contains a hydrophobic filter. If the filter media becomes wet, air flow will be stopped. The filter cartridge must then be replaced. Do not remove filter
media from filter cartridge.
NOTE– Filter cartridges are included with each disposable container. They are also available separately (7305D-636 6/pack or 7305D-635 12/pack).
changing Bacteria Filter (Reusable container):
1. Change bacteria filter if overflow occurs or every two months, whichever comes first.
2. Remove filter by disconnecting it from suction unit and lid assembly.
3. Replace with a clean DeVilbiss bacteria filter (non-sterile) and remount to suction unit and lid. Ensure that the clear side of the bacteria filter is toward elbow and
container when installing/re-installing. Do not reverse direction of filter. Additional filters (7305D-60812/pack) may be purchased from your authorized DeVilbiss
Healthcare provider.
NOTE– Do not substitute any other material for this bacteria filter. Substitution may lead to contamination or poor performance; use only DeVilbiss filters.
NOTE– Bacteria filter must be changed between patients.

TROUBLESHOOTING

NOTE– Your DeVilbiss Suction Unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO THE HOME MEDICAL
EQUIPMENT PROVIDER WHERE YOU PURCHASED IT, please take a few moments to check for these possible causes:
PROBLEM
Unit does not turn on, but green external power light is
illuminated.
Pump runs, but there is no vacuum.
Low vacuum.
Battery will not charge but charge light is illuminated.

SPEcIFIcATIONS/cLASSIFIcATIONS

Size
7305P & 7305D Series
weight
7305D Series
7305P Series
Electrical Requirements
Internal Rechargeable Battery
7305P Series
7305D Series
Vacuum Range
7305P & 7305D Series
Air Flow @ pump inlet:
container capacity
7305 D & P Series
7305 D & P Series - Optional
EN - 6
3/8
" tubing.
AcTION
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
3. If running from an internal battery, ensure that your unit has a battery installed.
4. If battery is installed, check that it is fully charged.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks or leaks.
3. Ensure that float shut-off is not activated.
4. Check for leaks or cracks in container assembly.
1. Use vacuum adjustment knob to increase vacuum level.
2. Check system for leaks.
3. Push vacuum adjustment knob in toward unit and then release.
1. Ensure that unit is equipped with an internal battery by contacting your DeVilbiss equipment provider.
2. Verify that charge light turns on.
3. Check electrical connections during charging.
4. Ensure wall outlet is live by plugging in a lamp.
H x w x D inches (cm)
9.0 x 7.0 x 8.0 (22.9 x 17.8 x 20.3)
lb. (kg)
3.8 (1.7)
6.3 (2.9)
100-240 V AC 50/60 Hz .75 A max; 12 V DC, 33 W max
Factory Equipped
Not Factory Equipped (provider-installed option)
80 to 550 mm Hg
27 LPM (free flow) typical (may be less when running from internal battery)
800 ml (cc) Disposable
1200 ml (cc) Reusable
A-704