Trouble Shooting

Switch oft the device before any procedure and unplug the power
cable from the socket.
PROBLEM
The device does not
work
The device does not
nebulise or nebulises
insufficiently.
18
CAUSE
The power cable has
not been correctly in-
serted into the power
socket
The medication has
not been in-serted in
the nebuliser.
The nebuliser has not
been prop-erly fitted.
The nebuliser nozzle
is clogged.
The air pipe is not
correctly con-nected
to the device
The air pipe is bent,
damaged or twisted
The air filter is dirty
SOLUTION
Correctly insert the power cable in
sockets.
Pour the right amount of medication
in the nebuliser.
Disassemble the nebuliser and reas-
semble it correctly as shown in the
connection diagram on the cover.
Disassemble the nebuliser, remove
the nozzle and perform cleaning
operations Failure to remove medi-
cation deposits from the nebuliser
affects its efficiency and operation
Strictly comply with the instructions
contained in the „Cleaning, Sanitiza-
tion and Disinfection" chapter.
Ensure the device's accessories are
properly connected to the air intake
(see connection diagram on the
cover).
Unwind the pipe and ensure it is not
crushed or punctured. lf necessary
replace it.
Replace the filter