Herunterladen Inhalt Inhalt Diese Seite drucken

Veroval compact BPU 22 Gebrauchsanleitung Seite 66

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Table 5
Guidance and manufacturer's declaration - electromagnetic Immunity
The Veroval
BPU 22 is intended for use in the electromagnetic environment specified below. The customer or the user of the Veroval
®
should assure that it is used in such an environment.
Test Frequency
(MHz)
Radiated RF
385
IEC61000-4-3
(Test speci-
fications for
450
ENCLOSURE
PORT IMMUNITY
to RF wireless
communications
equipment)
710
745
780
810
870
930
1720
1845
1970
Band a)
Service a)
(MHz)
380 –390
TETRA 400
430 –470
GMRS 460,
FRS 460
704 – 787
LTE Band
13, 17
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
1 700 – 1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Modulation b)
Modulation b)
(W)
Pulse
1,8
modulation b)
18 Hz
FM c)
2
± 5 kHz devi-
ation
1 kHz sine
Pulse
0,2
modulation b)
217 Hz
Pulse
2
modulation b)
18 Hz
Pulse
2
modulation b)
217 Hz
66
Distance
IMMUNITY TEST LEVEL
(m)
(V/m)
0,3
27
0,3
28
0,3
9
0,3
28
0,3
28
BPU 22
®