CALIBRATION
The unit was calibrated before leaving the factory. The suggested adjustment
cycle is one year. A calibration tool is available as accessory.
1
Remove the battery cover and make the following settings: weighting
filter A, measuring mode FAST, measuring range 60 to 110 dB.
2
Place the microphone of the SLM-500 as far as possible into the
calibrator's calibration chamber.
3
Switch on the calibrator and check the indication
on the SLM-500's display. If the indication varies
by more than 0.5 dB from 94 dB, the SLM-500
has to be readjusted. Use a screw driver to
adjust the trimming control until the display
indicates 94.0 dB
...............................................................................................................................................

TECHNICAL SPECIFICATIONS

Power supply: ..................... 4 x 1.5 V battery
............................................ or 9 V DC, 100 mA via power unit (max. 9 V DC)
Battery life:.......................... approx. 30 hrs.
Measuring ranges: .............. 30-80 dB, 40-90 dB, 50-100 dB, 60-110 dB,
............................................ 70-120 dB, 80-130 dB, 30-130 dB
Accuracy: ........................... ±1.5 dB at 94 dB, 1 kHz
Frequency range: ............... 31.5 Hz - 8.5 kHz
Level weighting: .................. A or C
Microphone: ........................ electret condenser microphone
AC output:........................... 0.707 V RMS, 600 ohms
DC output: .......................... 10 mV, 100 ohms
Dimensions (LxWxH):......... 145 x 110 x 60 mm
Weight: ............................... 0.6 kg
Accessory:
SLM calibration tool ............ No. 10360611