Herunterladen Inhalt Inhalt Diese Seite drucken

KERN MPE-HM Betriebsanleitung Seite 113

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

MPC_M / MPE_HM / MPE_PM-BA-e-1414
Place adjustment weight in the centre of the weighing pan
Wait until stability display „STABLE" appears
Confirm by pressing
be displayed.
The balance carries out a selftest, after that [Err19]
will be displayed and a signal will sound.
Switch off the balance
Take away adjustment weight
Turn on balance again, after the selftest the balance changes
into the weighing mode. Adjustment has now been completed
successfully.
(MPE) or
(MPC), [PASS] will
55

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Mpc-mMpe-pm

Inhaltsverzeichnis