Cleaning Instruction

Indication Unit
Prandtl tube
44
Measurement system and probe have to be
switched off and unplugged prior cleaning.
Fibres or other foreign bodies can be removed with
the following agents:
- Soapy water
- EDP plastic cleaner
- Window cleaning fluid (without ammonia)
- and for any stubborn deposits: pure benzene
These solutions are only to be applied to the
instrument with a fluff-free cloth.
Only carry out surface leaning! The instrument
is never to be immersed in the solution!
Is the prandtl tube dirty or blocked, we suggest the
following agents:
- running water
- soapy water
- pure benzene
Cleaning example:
1) Blow out Prandtl tube with compressed air very
carefully (max. pressure 0,5 bar)
2) Is the prandtl tube still blocked, rinse it out with
water (max. pressure 0,5 bar) or put the whole
tube in soapy water or pure benzene.
3) After cleaning it this way, rinse it out with
distilled water.
4) Rub prandtl tube carefully dry with a fluff-free
cloth and blow it out carefully.
5) Let it dry for approx. 30 min.