ManoAir500
By a long actuation of the ∆ ∆ ∆ ∆ p key the measuring instrument is
reset to the resolution of 0.05% of the upper limit for
pressure within +/- 10% of the upper limit.
Mean value measurements of the parameters ∆ ∆ ∆ ∆ p, Flow, Temp.
or Baro over a certain averaging time are made with the key
"Rep".
With the Mini2Logger connected and switched on, the Rep
measurement is automatically repeated by a long actuation of the
"Rep" key, and the data at the Mini2Logger is transferred between
two Rep measurements.
In the case of entries a point (comma) is set with the "Rep" key. A
point after the fourth place is not displayed.
Mean value measurements are initiated or interrupted with
the "Start / Stop" key.
It is also possible to define a particular mean value time for the
"Rep" key. With the key combination "Start / Stop" and "Rep" the
number of seconds for the mean value measurement is started in
the display. The "Start / Stop" key must be pressed again on
reaching the desired averaging time. This averaging time now
remains saved on the "Rep" key until it is either overwritten with a
new time or the instrument is switched off. (Standard averaging
time = 6 sec.)