Problem in Operation

1. If the meter does not display a message after inserting
a test strip:
PROBABLE CAUSE
Battery exhausted.
Battery incorrectly installed or
absent.
Test strip inserted upside down or
incompletely.
Defective meter.
2. If the test does not start after applying the sample:
PROBABLE CAUSE
Repeat the test using a new test strip with
Insufficient blood sample.
larger volume of blood sample.
Defective test strip.
Repeat the test with a new test strip.
Sample applied after automatically
Repeat the test with a new test strip.
shutoff (3 minutes after last user
Apply sample only when"
action).
appears on the display.
Defective meter.
Please contact the Customer Service.
106
WHAT TO DO
Replace the batteries.
Check that the battery is correctly
installed.
Insert the test strip correctly with the
contact bars end first and facing up.
Please contact the Customer Service.
WHAT TO DO
"flashing
3. If the control solution test result is out of range.

POSSIBLE CAUSE

Error in performing the test.
Do not shake the control solution
vial very well.
Expired or contaminated control
solution.
Control solution that is too warm or
too cold.
Test strip deterioration.
Meter malfunction.
WHAT TO DO
Read the instruction thoroughly and
repeat the test again.
Shake the control solution vigorously
and repeat the test again.
Check the expiration date or the
discarded date of the control solution.
Control solution, meter, and test strips
should come to room temperature
(20~25ºC/68~77ºF) before testing.
Repeat the test with a new test strip.
Contact the Customer Service.
107