STEP 3. Obtain Control Solution.
Shake the control solution vial well. Remove the cap. Squeeze the
vial, discard the first drop, and wipe off the remains on the tip of dis-
penser. Squeeze the vial again to get another drop of control solution
on your finger tip.
STEP 4. Apply Control Solution.
Move your finger (with control solution on it) to meet the absorbent
hole of the test strip and the drop will be automatically drawn into the
test strip. Remove your finger until the confirmation window is filled.
The meter begins to count down.
To avoid contaminating the control solution with the content of the test
strip, you have to place a drop of control solution on a clean surface or
on your fingertip. Do not directly apply control solution into a strip.
What
speaks?
Now testing.
82
STEP 5. Read and Compare the Result
After counting to 0, the test result of control solution is shown on the
screen. Compare this result with the range printed on the test strip
vial. It should fall within this range.
What
speaks?
Your blood glucose is 137 milligrams per deciliter.
Out-of-range results
If test results fall outside the range printed on the test strip vial, check
the section of "Problem in Operation" in troubleshooting guide and
repeat the test. If you continue to get out-of-range results, it means
that the system may not be working properly. Do NOT test your blood.
Contact the Customer Service for help.
Normal
109~165
83