Doing a Control Solution Test

STEP 1. Insert the Test Strip
Insert a test strip with contact bars end first and facing up, into the test
slot. The meter turns on automatically and displays the followings in
sequence:
→ "CHK" and "
"
→ "
" and ambient temperature
→ "
"and flashing " "
What
speaks?
Thank you for using this product.
↓ Test Strip Inserted.
↓ Today's date is Sunday, January 18th, 2008. The time is
10:08AM.
↓ The room temperature is 30.6 degrees Celsius.
↓ Please apply blood onto the test strip.
80
STEP 2. Press the
button
While the " " symbol appears on the display, press the M button and
then "CTL" will appear on the display. With the "CTL" sign on the dis-
play, the meter will not store your test result in memory. If you decide
not to perform a control solution test, press the M button again, and
the "CTL" sign will disappear.
What
speaks?
You are now in control solution mode.
WARNING
Contact bars must be inserted all the way into the meter or you may get an
inaccurate test result.
Every time you perform a control solution test, you must enter into the "CTL"
test mode so that the test result will not be stored in the meter memory.
Failure to do so will confuse the blood glucose test result with the control
solution test result in memory.