BEFORE TESTING

®
Checking with Monometer
Monometer
®
control solutions contain a known amount of glucose that
reacts with test strips. By comparing your control solution test results
with the expected range printed on the test strip vial label, it is able to
check that the meter and the test strips are working together as a sys-
tem and that you are performing the test correctly. It is very important
that you do this simple check routinely to make sure you get accurate
results.
How often the control solution test should be performed?
When you use this system to test your blood for the first time, prac-
tice the procedure using control solution. When you can do three
tests in a row that are within the expected range, you are ready to
test your blood.
For routinely check the meter and test strips, perform a single test
for each level of control solution at least once a week.
When the control solution test should be performed?
When you first get your Glucose Meter.
When you begin using a new vial of test strips.
Whenever you suspect that the meter or test strips are not working
properly.
When your blood glucose test results are not consistent with how
you feel, or when you think your results are not accurate.
When your test strips are exposed to extreme environmental conditions.
(See Storage section of this manual).
When you want to practice running the test.
If you drop the meter.
78
Control Solutions

Important Control Solution Information

®
Use only Monometer
control solutions.
Check the expiration date on the control solution vial. Do not use if
expired.
Control solution, meter, and test strips should come to room tempe-
rature (20~25°C/68~77°F) before testing.
Shake the vial, discard the first drop of control solution, and wipe off
the dispenser tip to ensure a good sample and an accurate result.
Use only for 90 days after first opening. Record the discard date (date
opened plus 90 days) on the control solution vial. Discard after 90
days.
Store the control solution tightly closed at temperatures below 30°C
(86°F). Do not refrigerate.
PLEASE NOTE
The control solution range printed on the test strip vial is for
trol solution only. It is used to test meter and test strip performance. It is not
recommended range for your blood glucose level.