Maximum (MAX) Minimum (MIN) Function
The MAX/ MIN mode allows the user to view only the lowest (MIN)
or highest (MAX) readings.
Press the MAX/ MIN button once and "MAX" appears on the display.
The LCD is now displaying the maximum reading of the selected
parameter in the display. The display will not change until a higher
reading is registered.
Press the MAX/ MIN button again and "MIN" appears on the display.
The LCD is now displaying the minimum reading of the selected
parameter in the display. The display will not change until a lower
reading is registered.
Press the MIN/MAX button for 3 seconds to exit the MIN/MAX
mode, and the "MAX" or "MIN" icon will disappear.
Auto Power OFF
Meter automatically shuts down after approx. 30 minutes of inactivity (Sleep
Mode). To override Auto Power Off function (Non-Sleep Mode), firstly press
the MAX/MIN button and hold, then press power button, after the display
appears, release both buttons. Icon "
unit will not auto power off now.
Replacing the battery
It is neccessary to replace the battery when the battery symbol
appears. Open the battery case on the rear side. Insert a new
alcaline battery. Connect the battery with the correct polarity with
the battery-clip. In order to avoid leaking batteries replace weak
batteries asap. In case of leaking batteries use gloves and goggles
for personal protection.

Technical information

Specifications
Function
Range & Resolution
Humidity
0.0 .. 100.0%RH
Temperature
-20.0 .. 60.0°C
(-4°F .. 140°F)
14
" will not shown on the LCD. The
Accuracy
±3.5%RH (20% .. 80%)
±5%RH
(0 .. 20% & 80 .. 100%)
±1°C(0°C .. 40°C)
±2°C(-20°C .. 0°C &
40°C .. 60°C)
Display:
Dual digital LCD display with backlight
Sensor Type:
A single chip relative humidity and temperature
multi sensor module comprising a calibrated
digital output
Response Time:
<15 seconds
(90% of final value in moving air)
Polarity:
Automatic, (-) negative polarity indication.
Low battery
The "
indication:
voltage drops below the operating level.
Operating
32 to 122°F (0 to 50°C);
Conditions:
< 80% RH non-condensing
Storage Conditions:
-40 to 185°F (-40 to 85o°C);
<99% RH non-condensing
Power:
One standard 9V, NEDA1604 or 6F22 battery.
Battery Life:
Approx. 48 hours
Dimensions/
146(L) x 54(W) x 34(H) mm;
Weight
138g (Including battery)

General provisions

Explanation of symbols
This product is CE compliant and thus fulfils the
necessary guide lines.
Disposal
Please do not despose of old electronic devices and
used batteries in house-hold waste.
To protect the environment, take them to your retail store
or to appropriate collection sites according to national or
local regulations.

Maintaining the unit

Clean it with a soft damp cloth. Do not use solvents or scouring agents.