Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Data - Wld-Tec SteriMax smart Bedienungsanleitung

Impfösen - sterilisator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Technical Data

Touch-Operation:
Status display:
IR-Sensor:
Thermocontrol:
Annealing tube
Quartz glass:
Safety features
Residual heat display:
Automatic unit switch off:
Temperature monitor:
Electrical
Power supply:
Power consumption:
Protection class:
Heat source:
Mechanical
Quartz glass annealing tube:
Casing:
Measurements (w x h x d):
Weight:
Information on radiation
Emission limit:
Licences
CE:
EU Guidelines:
Protected Design:
ON / OFF, selection for 2 users
sterilization timer 5 - 10 seconds
inoculation loop cooling timer 0 - 25 seconds
with audible signal (disengageable)
optical display of the sterilization progress
sensor range 8-stage adjustable
dynamic sterilization timer, reduces the
sterilization time according to the residual heat
750°C - 1000°C (1382°F - 1832°F) sterilization temperature
indicates a hot annealing tube
15, 30 oder 60 minutes
thermal circuit breaker
Art. No.: 4.001.000:
220 - 240 V 50/60 Hz 2.6 A
Art. No.: 4.002.000:
112 - 128 V 50/60 Hz 5.1 A
Art. No.: 4.003.000:
95 - 105 V 50/60 Hz 6 A
approx. 600 W / short time (5 - 10 seconds)
IP 20
IR-Halogen lamp
outer-Ø 19 mm, inner-Ø 16 mm, length 112 mm
heat resistant glass / stainless steel
UV and solvent resistant
110 x 170 x 180 mm
approx. 1800 g
The emission limit "risk free" is respected.
The device is assigned to the risk free group
according to EN 62471
EN 61326-1, EN 61010-1, EN 61010-2-010, EN 62471
2014/30/EU, 2014/35/EU, 2011/65/EU, 2006/25/EG
D DM/082 011
EN
31