Troubleshooting guide

• The displays do not light up after touching the On / Off symbol (1)
• Inoculating loops are not sterilized
• The sterilization process only starts every second time it is activated
• When the inoculating loop is inserted into the sterilization orifice (9a), the red
light does not light up on the sterilizer
• The unit gets hot, the red light turns off during sterilization
30
Switch on the device at the power switch (17).
Check if the power plug is correctly wired and make sure that it is
connected to a live wall socket (fit to the device).
Check the fuse; if the fuse is faulty, it can be replaced. To do this,
pull out the fuse holder (15). Replace the blown fuse with a new one
and push the fuse holder back into its socket.
The device is fitted with a (230V: 3,15 AmT; 100V / 120V: 6,30 AmT)
protection fuse. The factory provides a spare fuse in the reserve clip
in the fuse holder (15).
To replace a fuse, unplug the power cable (20) from the power
socket (16).
After switching off, the fans (11) run until the excess heat has
i
been dissipated.
Annealing tube is dirty. Please clean and dry it, or if necessary
replace it (see section 6.1).
Insertion mode is enabled. Please switch off (see section 4.6).
Clean the sensor window (8) or increase the sensor range
(see section 4.3 & 6.2).
The IR-Halogen lamp is faulty and must be replaced.
Error display "E2" (see section 8.2).
The thermal circuit breaker is switching the unit off.
Error display "E1" (see section 8.1).
Check the functioning of the fans (see section 7).