3.1 Starting the sterilization

Sterilization is started automatically
by inserting an instrument (such as
an inoculating loop with holder) into
the orifice of the annealing tube (9,
9a). Here, the user's hand must
be within the detection range of the
IR-Sensor (8). At the same time the
timer display (5) starts, and indicates
the remaining sterilization time. Hold
the instrument in the sterilization
orifice until the sterilization time has
expired. After expiration of the time
set, sterilization stops automatically.
The status display (S) shows the
course of the sterilization process.
Then remove the sterile instrument
from the annealing tube. The blue snowflake (7) illuminates and cooling down time
starts. After the cooling time (5) has expired, the device automatically switches back to
the sterilization time display (orange sun) (6).
Regardless of the cooling time, the sterilizer is ready for use again after 3 - 5 seconds.
If the buzzer is switched on, a warning tone sounds after the cooling time has
i
expired. If a new sterilization is started before the expiry of the cooling time, the
cooling down timer is stopped. If no cooling time is selected, the timer remains
enabled for the sterilization time.
The sterilization process can be stopped prematurely by briefly touching the
i
On / Off symbol (1) or the timer symbol (2).
Never hold fingers or other parts of the body on or in the sterilization orifice
(9a). Even some time after use, the sterilization orifice (9a) and the annealing
tube (9) are still hot. Danger of burns!
If the halogen lamp is illuminated, do not look directly into the sterilization
orifice (9a).

3.2 Thermocontrol

Depending on the frequency of sterilization, the thermocontrol makes use of the
residual heat, thus shortening the next sterilization automatically and dynamically
by up to 5 seconds.

4. Settings menu

This section describes the various settings of the device.

4.1 Settings

The settings menu is activated by briefly pressing the button (14)
of the device.
24