9 Error messages

Display
Description
OL or------
Weighing range has been exceeded (excess load)
------or Null
Below weighing range (underload)
Zero range exceeded
(on start-up or when pressing the
• Load on weighing pan
• Excess load, during zero setting of weighing scale
• Incorrect adjusting process
• Fault on load cell
Value outside the A/D converter range
• Damaged weighing cell
• Damaged electronics
Lack of possibility of zero point initiation
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
14
• Damaged / overloaded measuring cell
• The objects are located on the platform / have contact with
it
• Unremoved transportation protection
• Main Board damaged
key)
MPD-BAPS-e-1811