Assembly

Mounting periphery devices



Connect external temperature sensor Pt 1000

(optional accessory, see chapter "Accessories and spare parts")
Caution:
Risk of cable damage
The cable of temperature sensor
might get damaged by too high
temperatures.
Ensure the cable does not get in
contact with very hot materials.
Please use cable guide tube
delivered with holding system.
➜ Screw support rod into threaded hole at
left rear corner of device.
➜ Pass sensor holder over support rod.
➜ Adjust height of sensor holder and fi x
with knurled screw.
➜ Pass temperature sensor from top into
sensor holder.
➜ Clamp cable of temperature sensor into
three clamps of cable guide.
✓ Device is switched off.
➜ Connect plug to 4-pole socket with the
tongue facing down on back side of
device.
➜ Push locking ring towards device
and turn clockwise approx. 1/4, till it
audibly engages.
➜ Insert temperature sensor from above
into sensor holder.
Only connect and disconnect temperature sensor when device is
switched off.
If temperature sensor is connected or disconnected during heating
operation, the device is unable to measure and control temperature.
Support rod
Knurled screw
Sensor holder
Cable guide
Threaded hole
4-pole socket Pt 1000
29
Plug Pt 1000
Locking ring