Herunterladen Inhalt Inhalt Diese Seite drucken

Ecom Instruments MCAL 4200 Bedienungsanleitung Seite 87

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Read (mV)
Source (mV)
Maximum current output in voltage ranges is 1mA with an output impedance of
TC Type
J
K
T
E
R
S
B
C
XK
BP
L
U
N
All TC errors include CJC errors
CJC errors outside of 23 ± 5 °C is 0.05°C/°C (In °C add .2 for cold junction compensation error.)
86
Table 15: Thermocouple Measurement / Source
Range
-10.000mV - 75.000mV
-10.000mV - 75.000mV
Table 16: Thermocouple Read and Source
-210.0 ... 0.0
800.0 ... 1200.0
-200.0 ... 0.0
0.0 ... 1000.0
1000.0 ... 1372.0
-250.0 ... 0.0
-250.0 ... -100.0
-100.0 ... 1000.0
0.0 ... 1767.0
0.0 ... 1767.0
600.0 ... 800.0
800.0 ... 1000.0
1000.0 ... 1820.0
0.0 ... 1000.0
1000.0 ... 2316.0
-200.0 ... 800.0
800.0 ... 2500.0
-200.0 ... 0.0
-200.0 ... 0.0
-200.0 ... 0.0
0.0 ... 1300.0
Accuracy(% of reading ± digit)
0.02% ± 10
0.02% ± 10
Range (°C)
0.0 ... 800.0
0.0 ... 400.0
0.0 ... 800.0
0.0 ... 900.0
0.0 ... 600.0
< -
Accuracy (°C)
0.4
0.2
0.3
0.6
0.3
0.5
0.6
0.2
0.6
0.2
1.2
1.2
1.2
1.3
1.5
0.6
2.3
0.2
0.9
2.3
0.25
0.2
0.5
0.25
0.8
0.4

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Ecom Instruments MCAL 4200

Inhaltsverzeichnis