J. Questions and Answers
This Chapter provides the answers to frequently asked questions
Question
is shown in display
Detector switches itself
off or detector cannot be
switched on.
E32 is shown in display
E64 is shown in display
E96 is shown in display
E128 is shown in display
E160 is shown in display
E192 is shown in display
E224 is shown in display
If we were unable to answer your question. please contact your distributor or Testo
Customer Service. You will find contact details in the Representatives and Worldwide
Guarantee document or in Internet at www.testo.com .
Possible causes
Remaining life of rechargeable battery is
less than15min
Rechargeable battery empty
Calibration data incorrect
Sensor not connected
Calibration data incorrect and sensor not
connected
Parameter memory defective
Parameter memory defective and
calibration data incorrect
Parameter memory defective and sensor
not connected
Parameter memory defective and calibration data
incorrect and sensor not connected
J. Questions and Answers 15
Remedy
Recharge battery (Refer to
Recharging battery, P. 9).
Recharge battery (Refer to
Recharging battery, P. 9).
Send detector for repair to Customer
Service.
Send detector for repair to Customer
Service.
Send detector for repair to Customer
Service.
Send detector for repair to Customer
Service.
Send detector for repair to Customer
Service.
Send detector for repair to Customer
Service.
Send detector for repair to Customer
Service.