6.0

6.0 Cleaning

General instructions for cleaning and disinfecting the
exterior surfaces of the patient chair E2e
The chair must be separated from the mains
supply prior to cleaning.
Always observe the instructions for use by the
manufacturer of the disinfectants, above all,
concentration specifi cations and instructions.
The exterior surfaces of the patient chair E2e, except the
covering material, resist most of the common surface disin-
fectants.
The covering material is made of foam synthetic leather.
When cleaning and disinfecting this material pay attention
to the material compatibility. We recommend to clean the
upper part of the chair after each application.
Do not use any
disinfectants containing concentrated organic or
anorganic acids, since these may cause corrosion
damages,
disinfectants containing chloramides, phenol deri-
vates or anionic tensides, since these may cause
stress cracks in the plastic material used.

Recommended surface disinfectants

Disinfectant
TERRALIN
(Concentrate)
QUATOHEX
(Concentrate)
Incidin Plus
(Concentrate)
Pursept-A
(Disinfectant spray or
disinfectant cloths)
24
UK
Ingredients
Benzal conium chloride
Phenoxypropanole
Didecyl dimethyl
ammonium chloride
Benzal conium chloride
Bi-guanidinium acetate
Polymer biguanid
Cleaning agents
Glucoprotamin
Nonionic tensides
Solvents, complexing agents
Ethanol
Glyoxal
QAV
F
(in 100 g)
Manufacturer
20 g
Schülke & Mayr, Norderstedt
35 g
Braun, Melsungen
14 g
10 g
7.5 g
0.5 g
26.0 g
Henkel, Düsseldorf
38.9 g
Merz & Co., Frankfurt/M.
0.1 g
0.05 g
Page 25

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis