Switch point setting
Apply the operating voltage (+U
IN (pin 5)
OUT = 1
OUT = 0
Time
window
Duration [s]
Within the time windows A, B, C the output is switched depending on
the flow: output closed (OUT = 1) if flow ≥ SP / output open (OUT = 0) if
flow < SP.
If the flow rises or falls within the time windows A, B, C, the switching status
can change.
In the time windows D, E, F the output is used for feedback signals (→ table
below). It does not react to flow changes.
Time window Operation
A
Inititalisation of the setting operation
B
Confirmation of the initialisation
Switch point setting*
C
Signal U
results in SP [cm/s]:
Last switching status from C is maintained (= internal
D
monitoring).
Output signal is inverted (confirmation of the setting);
E
duration = setting time of the selected switch point).
Output signal is inverted again (= internal monitoring).
F
After this SP
*Accuracy: ± 1 s; factory setting: SP = 15 cm/s

Operation

After mounting and wiring check whether the unit operates correctly.
Recommended maintenance:
Check the sensor tip for build-up from time to time. Clean it with a
soft cloth. If necessary, build-up which adheres firmly (e.g. lime) can
be removed with a common vinegar cleansing agent.
) to pin 5 for the specified time.
B
at pin 5 [s]:
B
min 10
is active.
NEW
5
10
15
20 ... 55
15
20 ... 55
60
65
60
max
7